22 Lý Tự Trọng, Minh Khai, Hồng Bàng, Hải Phòng.
itvc.haiphong@itvc-global.com
Chi tiết sản phẩm
THÔNG SỐ KỸ THUẬT:
Hãng sản xuất: Fluke
Model: P3011-KPA-P (kèm theo bơm)
Hoạt động chân không
Kiểu PCU đơn
Giải áp suất: 3 đến 100 kPa
Độ chính xác: ±0.015% của giá trị đo
Ngõ ra : 1/8, 1/4, 3/8 và ½ NPT hoặc BSP (option 5540)
Kích thước: 17.5 W x 12 D x 8.5 H
Option:
- 5540 Metric and 5541 BSP Adapter Sets:
+ P5540 Metric: M14 x 1.5, M20 x 1.5, 1/8 and 1/4 BSP
+ P5541 BSP: 1/8, 1/4, 3/8 and 1/2 BSP
+ P5542 NPT: 1/8, 1/4, 3/8 and 1/2 NPT
- P5531 Dirt/Moisture Trap: Pressure range 0 to 500 psi (35 bar)
P5532: Dirt/Moisture Trap Pressure range 0 to 3,000 psi (210 bar)
P5543: Angle Adapter:
P5544: Two Gauge Stand
P5551 Pointer Remover/Punch
Accuracy |
|
Accuracy* |
±0.015% of reading (±0.008% optional) *Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. |
Mass and PCU Materials of Construction |
|
Standard weight material |
Series 3 non-magnetic austenitic stainless steel |
Weight density |
7.8 g/cm³ |
Optional fractional weights |
Solution heat treated aluminum |
Weight density |
2.7 g/cm³ |
Models P3011, P3012, P3013, P3022, P3023, P3025 (V) |
Piston Material: Ceramic |
Models P3014, P3015, P3025 (P) P3031, P3032 (P & V) |
Piston Material: Tungsten carbide |
Model P3016 |
Piston Material: Tungsten carbide |
General |
|
Test port adaptors |
1/8, 1/4, 3/8 and ½ NPT or BSP |
Instrument weight |
24 lbs |
Instrument size |
17.5 W x 12 D x 8.5 H |
Mass set weight (typical) |
65 lbs |
Minimum Standard Weight Increments |
|
P3012, P3022 Pressure |
1 inH2O, (5 mbar) |
P3013, P3023 Pressure |
2 inH2O, (10 mbar) |
P3014, P3015, P3025 Pressure |
1 psi, (0.1bar) |
P3016, P3031, P3032 |
1 psi, (0.1bar) |
P3011, P3022, P3023, P3025 Vacuum |
0.2 inHg (10 mbar) |
Optional Fractional Weights |
|
P3014, P3015, P3025 Pressure |
0.1 psi (0.01 bar) |
Built-in Hand Pumps |
|
Pressure mode |
Max pressure 300 psi (20 bar) |
Vacuum mode |
To 90% vacuum |
Operating Fluid for Liquid Lubricated Models |
|
55-655 |
mineral oil (standard) |
Krytox |
for oxygen-safe applications (optional) |
Options |
|
PressCal Software |
Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments. |
Krytox |
As operating fluid for liquid lubricated gas operated models. |