22 Lý Tự Trọng, Minh Khai, Hồng Bàng, Hải Phòng.

itvc.haiphong@itvc-global.com

Thiết bị đo đa năng KYORITSU 6010B, K6010B

(Click vào để xem ảnh lớn)

Thiết bị đo đa năng KYORITSU 6010B, K6010B

Giá: Liên hệ Lượt xem: 0 Hàng mới

+ Chỉ thị số
+ Đo thông mạch, đo mạch vòng, đo điện trở đất...
+ Phạm vi đo: 20/200MΩ /: 20/2000Ω ;  500/1000V
Bảo hành : 12 tháng
Hãng sản xuất : Kyoritsu - Nhật
Xuất xứ : Nhật

Chi tiết sản phẩm

Continuity:
20/200Ω
200mA Test Current
Insulation:
500/1000V, 20/200MΩ
1mA Rated Test Current
Loop Impedance:
20/2000Ω
15mA Test Current at 2000Ω (No RCD Trip)
RCD:
10/30/100/300/500mA
x1/2, x1, Fast Trip(40ms), DC, Auto Ramp
Test of a large kind of RCD's:
Standard, Selective, AC and A
(DC sensitive breakers)
Uc(Contact Voltage): 100V
Designed to IEC 61010-1, IEC 61557
Data Memory:300 measured results
The stored data can be transferred to PC via Optical Adaptor Model 8212-USB or 8212-RS232C.
6010B
CE
 
Continuity Testing
Measuring Range 20/200Ω (Auto-Ranging)
Open Circuit Voltage > 6V
Short Circuit Current > 200mA
Accuracy ±(3%rdg + 3dgt)
Insulation Resistance Testing
Measuring Range 20/200MΩ (Auto-Ranging)
Test Voltage 500/1000V
Open Circuit Voltage +20% - 0%
Rated Current > 1mA
Accuracy ± (3%rdg + 3dgt)
Loop Impedance Testing
Impedance Range 20/2000Ω
Rated Voltage 230V + 10% - 15%(50Hz)
Normal Test Current 20Ω:25A/10ms
2000Ω:15mA/350ms max.
Accuracy ±(3%rdg + 8dgt)
RCD Testing
Test Current
(Test Current Duration)
x1/2, x1 10, 30, 100, 300, 500mA (2000ms)
Fast 150mA (50ms)
DC 10, 30, 100, 300mA (2000ms), 500mA (200ms)
Auto Ramp Goes up by 10% from 20% to 110% of IΔn. 300ms×10
Rated Voltage 230V + 10% - 15% 50Hz
Accuracy Test Current x1/2: -8%-2%
x1, FAST: +2%+8%
DC: ±10%
Auto Ramp: ±4%
Trip Time ±(1%rdg + 3dgt)
Uc Testing
Measuring Range 100V
Rated Voltage 230V + 10% - 15% (50Hz)
Test Current 5mA at IΔn = 10mA
15mA at IΔn = 30/100mA
150mA at IΔn = 300/500mA
Accuracy +5%+15%rdg, ±8dgt
General
Safety Standard IEC 61010-1 CAT.III 300V Pollution Degree 2
IEC 61557-1, 2, 3, 4, 6, 10, IEC 60529 (IP40)
Power Source R6 or LR6 × 8
Dimensions 175(L) × 115(W) × 86(D)mm
Weight 840g approx.
Accessories 7122B (Test leads) 
KAMP10
* (Test Lead with IEC Connector)
Pouch for Test leads
Shoulder Strap
Shoulder Pad
Instruction Manual
Optionals 7133B (Distribution board test leads)
8212-RS232C (RS232C adaptor with "KEW Report (Software)")
8212-USB (USB adaptor with "KEW Report (Software)")
* KAMP10 (AU): Australian plug
KAMP10 (UK): British plug (13A)
KAMP10 (EU): European SHUKO plug
KAMP10 (SA): South African plug
 
Gọi ngay: 0914.564.579
SMS: 0914.564.579 Chat Zalo Chat qua Messenger