Measurement Accuracy |
|
Voltage DC |
|
30.000 V |
0.02%+ 2 counts (upper display) |
30.000 V |
0.02%+ 2 counts (lower display) |
100.00 mV |
0.02%+ 2 counts |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
Current DC |
|
24.000 mA: |
0.02%+ 2 counts |
Resistance |
|
0.0 to 400.0 Ω |
0.1Ω (4-wire), 0.15Ω (2- and 3-wire) |
401 to 1500 Ω |
0.5Ω (4-wire), 1Ω (2- and 3-wire) |
1500 to 3200 Ω |
1Ω (4-wire), 1.5Ω (2- and 3-wire) |
Frequency |
|
2.0 to 1000.0 CPM |
0.05 % + 1 count |
1.0 to 1100.0 Hz |
0.05 % + 1 count |
1.00 to 10.00 kHz |
0.05 % + 1 count |
Sensitivity |
1 V peak-to-peak-minimum |
Pressure |
|
Accuracy |
from 0.025% of range using any of 8 intrinsic safe pressure modules. (for detailed specifications refer to pressure modules in options and accessories) |
Source Accuracy |
|
Voltage DC |
|
100.00 mV |
0.02%+2 counts |
10.000 V |
0.02%+2 counts |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
Current DC |
|
24.000 mA (Source) |
0.02%+ 2 counts |
24.000 mA (Simulate) |
0.02%+ 2 counts |
Resistance |
|
15.0 to 400.0 Ω |
0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
401 to 1500 Ω |
0.5 Ω(excitation current 0.05 to 0.8 mA) |
1500 to 3200 Ω |
1 Ω(excitation current 0.05 to 0.4 mA) |
Frequency |
|
2.0 to 1000.0 CPM |
0.05% |
1.0 to 1100.0 Hz |
0.05% |
1.00 to 10.00 kHz |
0.25% |
Waveform: |
5 V p-p squarewave, -0.1 V offset |
RTDs and Thermocouples |
|
Measure Accuracy: |
|
NI-120 |
0.2 °C |
Source Accuracy: |
|
NI-120 |
0.2 °C |
Specifications |
|
Ramp Functions |
|
Source functions: |
Voltage, current, resistance, frequency, temperature |
Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
Loop Power Function |
|
Voltage: |
24 V |
Accuracy: |
10% |
Maximum current: |
22 mA, short circuit protected |
Step Functions |
|
Source functions: |
Voltage, current, resistance, frequency, temperature |
Steps |
25% of range, 100% of range |
Environmental Specifications |
|
Operating Temperature |
-10 °C to +55 °C |
Storage Temperature |
-20 °C to 71 °C |
Operating Altitude |
3000 m |
Safety Specifications |
|
Agency Approvals |
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
Mechanical & General Specifications |
|
Size |
130 x 236 x 61 mm |
Weight |
0.65 kg |
Batteries |
4 AA alkaline batteries |
Battery Replacement |
Separate battery compartment, accessable without breaking calibration seal |
Side port connections |
Pressure module connector, also used for remote real-time programming |
Manufacturer : Fluke - USA
Description
“Simply Powerful!”
Take along the new Fluke 725 Multifunction Process Calibrator, and you're equipped to test and calibrate almost any process parameter. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.
The 725's split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725 has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.
From the moment you pick it up, the Fluke 725 is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.
The Fluke 725 Multifunction Process Calibrator is a powerful, yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter.